Troubleshooting
Message
Calibration invalid. Please
perform calibration.
Measurement failed! Please
repeat. Ambient light.
Instrument temperature too
high!
Instrument temperature too
low!
Memory full! Please delete
stored measurements.
Sample to bright! Please
switch off M-indices ...
Gloss and temperature out
of range.
Temperature out of range.
Gloss out of range.
88
Solution
Perform calibration the external white calibration standard.
Appears if an error occurs during measurement.
Make sure to completely cover measurement aperture.
Make sure to hold the instrument stable during measurement.
Repeat measurement.
Instrument temperature is > 45°C.
Allow the instrument to cool down.
Instrument temperature is < 5°C.
Allow the instrument to warm up.
Delete stored measurements.
Disable the jetness indices and repeat.
Allow instrument to cool down and repeat.
Allow instrument to cool down and repeat.
Repeat measurement on a different spot.