If the diagnosis passes, you can use the instruments to measure. If the diagnosis fails, following
message appears: "Please clean jetness test standard or calibrate on external gray calibration
standard".
Clean the test standard and repeat. If diagnosis still fails, calibrate the instrument on the gray
74
standard, see
Calibration on Gray
.
6.7.4

Calibration on White

Calibration on white is required if the diagnosis for color or gloss fails. Click on Calibrate on
white standard. The instrument will guide you through the calibration procedure.