color2view
If the calibration passes, repeat the diagnosis, see
fails, following message appears: "Please clean standard / Color, Fluorescence, Gloss out of
range".
Clean the standards and repeat. If calibration still fails, contact the BYK-Gardner service team.
NOTICE
Perform calibration only, if required in order to ensure correct status of your calibration
standard.
6.7.5
Calibration on Gray
Calibration on gray is required if the diagnosis for jetness fails. Click on Calibrate on white
standard. The instrument will guide you through the calibration procedure.
If the calibration passes, repeat the diagnosis, see
following message appears: "Please clean standard / Color out of range".
74
Diagnosis Color / Gloss
Diagnosis for Jetness
BYK-Gardner GmbH
72
. If the calibration
72
. If the calibration fails,