Use
Difference Measurement
performance. If the differences are within your tolerances you can continue to measure with your
instrument. Otherwise calibrate your instrument.
The option Add to standard list will be inactive next time when this standard has already been
added to the instrument memory. To delete the item from the memory use the function
48
Measurements
>
Delete Standard
NOTICE
Every difference measurement is stored in the instrument. Transfer the measurement
data to "smart-chart" to document each test case.
6.7.7

Lens Cleaning

In order clean the lenses of the instrument select Open standard paddle. The shutter opens to
allow access to the measuring optics.
When cleaning was done select Close standard paddle. The shutter closes to protect the
measuring optics.