BYK-Gardner GmbH
6.7
Calibration
The system delivery comprises test and calibration standards, see
21
Standards
. These tiles can be used to check if your instrument is measuring correctly. Each
standard tile comes with a dedicated serial number matching your instrument. In the
configuration menu select Calibration.
The instrument calibration comprises following functions:
·
Calibration Information
·
Diagnosis Color / Gloss
·
Diagnosis for Jetness
·
Calibration on White
·
Calibration on Gray
·
Saving Standards
·
76
Lens Cleaning
When using these tiles take care for correct alignment of instrument.
6.7.1
Calibration Information
Use the information screen to check the current status of diagnosis and calibration.
The screen shows the details for instrument status and for calibration and test standards.