Operational Capabilities; Table 8.4: Gas Supply Specifications; Table 8.5: Operational Capabilities - Ox-Tran 2/28 Model H - Ametek mocon OX-TRAN 2/28 H Operator's Manual

Oxygen transmission rate system
Table of Contents

Advertisement

Specifications
Gas Supply Requirements
Gas Composition
Supply Pressure, Nominal
Supply Pressure, Maximum
Carrier Gas
Supply Pressure, Minimum
Flow Rate, Nominal, Test Active, Maximum
Flow Rate, Nominal, Test Active, Minimum
Gas Composition
Supply Pressure, Nominal
Test Gas
Supply Pressure, Maximum
Supply Pressure, Minimum
Flow Rate, Nominal, Test Active

Operational Capabilities

Temperature Control Range
Time to Reach a Heated Temp (from ambient)
Test Range, Film, 50 cm
Test Range, Package
Resolution
Repeatability
8-2

Table 8.4: Gas Supply Specifications

2

Table 8.5: Operational Capabilities - Ox-Tran 2/28 Model H

OX-TRAN Model 2/28 Operator's Manual
Nitrogen/Hydrogen Mix (98% / 2%)
25 PSI, (1.72 Bar), (172 kPa)
28 PSI, (1.93 Bar), (193 kPa)
20 PSI, (1.38 Bar), (138 kPa)
111.5 cc/minute
58.5 cc/minute
100% Oxygen (purity 99.9%)
25 PSI, (1.72 Bar), (172 kPa)
32 PSI, (2.2 Bar), (221 kPa)
20 PSI, (1.38 Bar), (138 kPa)
20 cc/minute
20 °C to 60 °C ± 0.2 °C
Approximately 60 minutes to reach 99% of a 40 °C setting
0.05 to 200 cc/(m
2
0.0032 to 12.9 cc/(100in
0.0003 to 1.0 cc/(pkg • day)
0.0001 cc/(m
• day)
2
0.000006 cc/(100in
0.0000005 cc/(pkg • day)
± 0.02 cc/(m
• day)
2
± 0.0013 cc/(100in
± 0.0001 cc/(pkg • day)
Revision J
• day)
• day)
2
• day)
2
• day)
2
or 1% of value, whichever is greater