Mounting A Sample In The Top Test Cell Position; Figure 5-2: Mounting A Film Sample (Top Position) - Ametek mocon OX-TRAN 2/28 H Operator's Manual

Oxygen transmission rate system
Table of Contents

Advertisement

OX-TRAN Model 2/28 Operator's Manual

Mounting a Sample in the top Test Cell position

1.
Place the partially assembled Test Cell on a flat surface with the top mounting surface up.
2.
Place the film on the greased sealing surface, remove any wrinkles as necessary.
3.
Verify the inner and outer O-Rings are properly seated in the top Test Gas Plate.
4.
Install the top Test Gas plate on the Film Mounting Plate.
Note: Inspect the gap between the top and bottom Test Gas Plates and the Film Mounting plate. If the gap
is not even on all four sides one of the O-Rings may not be properly seated.
MOCON, Inc.

Figure 5-2: Mounting a Film Sample (Top Position)

Revision J
Testing Film Samples
5-5

Advertisement

Table of Contents
loading

Table of Contents