Mounting Samples in the Test Cell

Follow the instructions below to mount a film sample in the Test Cell:
Note: For information on cutting and preparing the film samples refer to the sections on "Sample Size",
"Using Masks" and "Orienting the Sample" in this chapter.
1.
Remove the Test Cell from the instrument (see Loading and Unloading the Test Cell).
2.
Separate the top and bottom parts of the cell from the Film Mounting Plate (middle part).
3.
Remove the old film samples from both sides of the Film Mounting Plate.
4.
Clean the film sealing surfaces of the cell to remove the old grease.
5.
Apply new Apiezon grease to the sealing surfaces on the Film Mounting Plate.
6.
Inspect the TruSeal flushing ring and ports, remove any excess grease.

Mounting a Sample in the bottom Test Cell position

1.
Place the Film Mounting Plate on a clean flat surface with the bottom mounting surface up.
2.
Place the film on the greased sealing surface, remove any wrinkles as necessary.
3.
Verify the inner and outer O-Rings are properly seated in the bottom Test Gas Plate.
4.
Install the bottom Test Gas plate on the Film Mounting Plate.
5-4

Figure 5-1: Mounting a Film Sample (Bottom Position)

OX-TRAN Model 2/28 Operator's Manual
Revision J