Minor Repair; General Information; Printer Replacement - Bayer HealthCare CLINITEK 500 Operating Manual

Urine chemistry analyzer
Hide thumbs Also See for CLINITEK 500:
Table of Contents

Advertisement

MINOR REPAIR

General Information

The CLINITEK
®
500 Urine Chemistry Analyzer is a self-
contained instrument that requires very little maintenance.
With proper care and use, the instrument should oper-
ate reliably with a minimum of operator attention. The
only minor repair that can be performed is the replace-
ment of the printer, should this ever become necessary.
In addition, the touch screen can be recalibrated, if
needed. This section is provided as an aid in performing
these procedures; for any other repairs, refer to Section
8, TROUBLESHOOTING AND SERVICE, for instructions
on obtaining service for your instrument.

Printer Replacement

Tools Required:
None
Parts Required:
Printer (Part No. 40451012)
Procedure:
1. WARNING: TURN THE INSTRUMENT OFF AND
REMOVE THE AC ELECTRICAL CORD FROM THE
OUTLET.
2. Remove the cover on the internal printer by pressing
in on the lever on the back of the printer cover and
lifting up on the cover (Figure 7-1).
Revised 9/99
Figure 7-1
3. Remove the roll of paper from the printer by tearing
the paper between the roll and the printer, then gen-
tly pull the paper through the printer in its normal
direction of travel or rotate the wheel on the right side
to advance the paper.
4. To remove the printer:
4.
a. Carefully remove the clear plastic shield that is
adhered to the front of the printer. Be careful not
to tear the adhesive pads.
4.
b. The printer hooks under a ledge at the front and
snaps into position at the rear with a small lever
(Figure 7-2). Press out on the lever and lift up on
the printer. Raise the module out of its cavity by
lifting up and back.
Section 7
7.1