SELECTING YOUR OPTIONS
H. Setup Menu #7
When all options have been selected on the sixth menu,
touch
to progress to the next menu in the Setup
Routine. The menu is displayed as, for example:
Set options.
Edit flagged results-OFF
Enter sample IDs-OFF
Tech ID-OFF
1. Edit flagged results:
1.
Results that have been flagged as positive and select-
ed for the confirmatory report can be edited (for exam-
ple, if confirmatory testing has been performed). Touch
the cycle key to turn the option ON or OFF, as desired.
This option can be selected only if "Mark Positives"
(Step E-2 previously) is ON.
2. Enter sample IDs:
2.
An identification number can be entered for each spec-
imen, either as part of a load list or immediately prior
to testing the specimen, if this option is set to ON.
Touch the cycle key to turn the option ON or OFF, as
desired.
3. Tech ID:
3.
The technician who is performing the testing can be
identified on the test results, if desired, by setting this
option to ON. The Tech ID can appear on control results
only or on both patient and control results. It is shown
on the Ready/Run screen and is easily changed when
necessary (see "Getting Ready to Run," Step 4, in
Section 4). Touch the cycle key to select the desired
option (OFF; ON, control results only; ON, both patient
and control results).
3.12
I. Setup Menu #8
When all options have been selected on the seventh
menu, touch
to progress to the next Setup menu. The
menu is displayed as, for example:
Set options.
Computer port options
Bar code reader options
Password for setup-OFF
Set or reset password
1. Computer port options:
1.
If you are sending results to a host computer or LIS
(Laboratory Information System), the interface param-
eters must be specified. Touch the Computer port
options key to display a new menu.
1.
a. The first three menu options feature cycle keys to
select the correct parameters. The following table
shows the possible selections for these options:
Port:
Baud:
1200
ON
2400
OFF
4800
9600
19200
Data, Parity:
8/NONE
7/EVEN
7/ODD
7/NONE
Revised 5/01