INSTALLATION
Figure 2-6
7. Hold the program card with the label facing forward
and the arrows pointing in and up. Insert the card into
the card receptacle (Figure 2-7) and press it in firmly
until the button above the receptacle is pushed out.
When properly inserted, the edge of the card will be
flush with the side of the instrument.
Figure 2-7
2.4
8. Install a roll of printer paper and re-install the printer
cover as follows:
a. Notice the large tab on the back side of the instru-
ment that secures the cover in place (Figure 2-8).
Press in firmly on the bottom edge of the tab and
lift the cover off.
Figure 2-8
8.
b. Obtain a new roll of paper; unroll several inches and
trim the end into a long "V". Hold the roll just above
the printer, with the paper unrolling from under-
neath. Feed the end of the paper under the roller,
then rotate the paper advance wheel in a clockwise
direction (toward the back) until several inches of
paper are exposed above the printer (Figure 2-9).
Revised 9/99