Bayer HealthCare CLINITEK 500 Operating Manual page 35

Urine chemistry analyzer
Hide thumbs Also See for CLINITEK 500:
Table of Contents

Advertisement

1. a.
The port must be ON in order to transfer results to
a computer. Refer to the specifications accompa-
nying the computer for information on the required
parameters for Baud, Data, and Parity.
1. a.
NOTE: The computer port can also be turned ON
or OFF through the first Setup menu, which is
unrestricted regardless of whether password pro-
tection is being used (see Step I-3 later). If use of
the port is changed (from ON to OFF or vice versa)
in one menu, the selection is automatically changed
in the other menu.
1.
b. The fourth option on the Computer port options
menu allows selection of the Output Format for the
results sent to a computer. A new menu is dis-
played with the following options and their possi-
ble selections:
Output Format:
Checksum:*
CCS
ON
CT200
OFF
CT200
*Not available in CCS format.
1. b.
The output format you select will depend upon
whether you already have a software interface pro-
gram that allows the transfer of results from either
a CLINITEK
®
200+ or a CLINITEK
Chemistry Analyzer to a host computer or LIS.
When one of these formats is selected ("CT200+"
or "CT200"), the data from the CLINITEK 500
instrument will be transmitted in the same format
as that sent by the selected instrument (the print-
ed results, however, do not mimic the selected
instrument).
1. b.
Alternatively, results can be transmitted in the
CLINITEK 500 format by selecting "CCS." The
parameters for this format are available from your
Bayer representative or office.
Revised 5/01
1. b.
1. b.
1. b.
Handshake:*
ON
OFF
1.
c. Select the correct parameters for all options, then
2. Bar code reader options:
2.
An optional Handheld Bar Code Reader is available for
use with the CLINITEK 500 Analyzer. This reader can
®
200 Urine
be used to scan barcoded identification labels on each
specimen cup or tube, rather than entering the speci-
men ID manually. The Analyzer software must be con-
figured to the appropriate parameters for the barcoded
labels being used. Touching the Bar code reader
options key will change the display to a new menu.
2.
a. Label: The Handheld Reader can accept any of four
If either CT200+ or CT200 is selected as the out-
put format, the use of checksum and handshake
must also be specified (refer to your computer
specification for the requirements). It is also very
important that you check the selections made on
your CLINITEK 200+ or CLINITEK 200 instrument
and enter the parameters identically in the
CLINITEK 500 Analyzer (e.g., ID/color/clarity/mark
positives ON or OFF). This will help ensure that the
data is transferred in the same format as your soft-
ware program currently is written to accept.
NOTE 1: Two stop bits are always transmitted in
the CT200+ and CT200 formats; one stop bit is
transmitted in the CCS format.
NOTE 2: If the "CT200" format is selected, results
obtained using any of the MULTISTIX PRO Reagent
Strips will not be sent to the computer. This is
because the new tests on the MULTISTIX PRO
Strips are not recognized by the CLINITEK 200
Analyzer (and therefore by its corresponding soft-
ware interface program).
touch
as needed to return to the previous Setup
menu.
different bar code formats. The format being used
can be selected using the Label cycle key, or you
3.13