Bayer HealthCare CLINITEK 500 Operating Manual page 74

Urine chemistry analyzer
Hide thumbs Also See for CLINITEK 500:
Table of Contents

Advertisement

COMPUTER/PRINTER INTERFACE
Pin Assignments for Interface Cable — Serial Port
Pin
Signal
Number
Name
1
CHAS GND
2
TXD
3
RXD
4
RTS
5
CTS
6
DSR
7
SIG GND
20
DTR
(All other pins are unused.)
Hardware Handshaking:
RXD
Receive Data
TXD
Transmit Data
DTR
Data Terminal Ready
DSR
Data Set Ready
RTS
Request To Send
CTS
Clear To Send
The following signal lines are NOT implemented:
DCD
Data Carrier Detect
RNG
Ring Indicator
CPI.2
Function
Protective Ground
Transmit Data
Receive Data
Request To Send
Clear To Send
Data Set Ready
Signal Ground
Data Terminal Ready
This input receives control characters for software handshaking and data for IDs.
This output sends test data, control characters, and instrument information.
This signal is on whenever the instrument I/O is configured for a computer and the
instrument is on.
The computer must raise this line whenever it is ready to receive data. If not supplied
by the computer, pin 6 must be jumpered to pin 20.
This output line, when high, indicates to the computer that it may send data.
This input is checked before sending each character and, if high, the next character
is sent. If not supplied by the computer, pin 5 must be jumpered to pin 4.
Pin 8
Pin 22
Table CPI-1
Signal
Type
Source
Ground
Data
CLINITEK 500
Data
Computer
Control
CLINITEK 500
Control
Computer
Control
Computer
Ground
Control
CLINITEK 500
N/A
N/A
Revised 5/01