4. Install the moving table as follows:
4.
a. Hold the table with the small rectangular tab fac-
ing to the back.
4.
b. Align the two grooves on the bottom of the table
with the edges of the platform on which the table
rests (Figure 2-4).
Figure 2-4
2.
c. Gently push the table in as far as it will go. It must
be pushed past a detent in order to be correctly in
position.
5. Next, install the fixed platform. (The holddown should
already be securely installed onto the platform. If it is
not installed, or is loose, refer to Section 5, "Daily
Cleaning," Step 11-a, for directions on installing the
holddown.) Align the two grooves on the bottom of
the platform with the arms extending from the instru-
ment, as shown in Figure 2-5. (The ledges on the left
and right sides of the holddown align just outside the
read area cover, and the top edge of the platform aligns
just under the cover.) Gently push the platform in as
far as it will go. (It must be pushed past a slight detent
to be correctly positioned.)
5.
Revised 9/99
CAUTION: If the platform does not push in at least
halfway with only very gentle pressure, do not force
it! Ensure that the moving table is correctly positioned
and attempt to reinstall the platform.
5.
NOTE: With the initial installation, you may need to
use firm pressure to push the platform the final
inch (1.3 cm). If the platform is not fully seated, or
1
2
if it is slightly crooked, the strips may jam as they are
pushed along the platform.
Figure 2-5
6. Hold the push bar by its indented end and, with this
end slightly upward, insert the peg on the other end
of the bar into the hole in the pusher mechanism
(Figure 2-6). Lower the push bar into place.
2.3