Function Inspection - Rigol MSO7000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO7000 Series:
Table of Contents

Advertisement

Chapter 1 Quick Start

Function Inspection

1.
Press Default on the front panel, then a prompt message "Restore default?" is
displayed. Press OK or tap OK to restore the instrument to its factory default
settings.
2.
Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 1-10 below.
3.
Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-10.
4.
Set the probe attenuation to 10X, and then press AUTO.
5.
Observe the waveform on the display. In normal condition, the square waveform
as shown in Figure 1-11 should be displayed.
6.
Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation" introduced in the next section.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the
metallic part of the probe when the probe is connected to high voltage
source.
MSO7000/DS7000 User Guide
Figure 1-10 To Use the Compensation Signal
Figure 1-11 Square Waveform Signal
Compensation Signal Output Terminal
Ground Terminal