Chapter 11 Protocol Decoding
Set the label display
Press Label to enable or disable the label display of the CAN decoding bus.
When enabled, the bus label "CAN" will be displayed at the upper-left side
of the bus (when the bus display is enabled).
Event table
Press Event Table to enter the event table setting menu. The event table
displays the detailed decoding information in time order in the form of a
table, so as to better observe the long decoded data. The decoding
information includes the decoded data, the corresponding line number, time,
frame ID, data length code (DLC), cyclic redundancy check (CRC), and
Acknowledgement (ACK) information.
Open or close the event table: Press Event Table continuously to
enable or disable the event table. When you enable the event table,
the following figure is displayed, as shown in Figure 11-11. You can
also enable the touch screen to tap the icon
corner of the table to close the event table. Besides, when the
operating status is "STOP", you can also rotate the multifunction knob
to select a specified line and view the corresponding decoding
information.
Line Number
Event Table
CAN Bus
Note:
MSO7000/DS7000 User Guide
Time Frame ID DLC Decoded Data
Figure 11-11 CAN Decoding Event Table
When you adjust the horizontal time base, the waveform displayed
on the screen will also change, and the total number of lines
containing the decoding information in the event table will also be
changed.
The displayed decoded data information in the bus is related to the