Rigol MSO7000 Series User Manual page 242

Digital oscilloscope
Hide thumbs Also See for MSO7000 Series:
Table of Contents

Advertisement

RIGOL
decoding
function
automatically). For analog channels, the threshold level settings will also be
copied.
3. Set the source and the threshold
Press Source and rotate the multifunction knob
source channel, and then press down the knob to select it. You can also
press Source continuously or enable the touch screen to select it. The
available channels include CH1-CH4 and D0-D15.
When you select the analog channel (CH1-CH4), press Threshold, then
rotate the multifunction knob
the threshold of the source channel. When you modify the threshold of the
clock channel, a dotted line displaying the current threshold level is
displayed on the screen. The dotted line disappears in about 2 s after you
stop modifying the threshold.
4. Signal setting
Press Settings to enter the signal setting menu.
Set baud rate
Three setting methods are available for baud rate:
Press Baud Rate, then set the user-defined baud rate with the
displayed numeric keypad.
Press Baud Rate, then select the preset baud rate from the available
options. The available baud rates include 2.4 kbps, 4.8 kbps, 9.6 kbps,
19.2 kbps, and etc.
Press Baud Rate, then rotate the multifunction knob
baud rate at a small step.
Set the parity bit
Press Parity bit continuously to select whether the data contains the parity
bit.
Set the protocol version
Press Version and rotate the multifunction knob
version that matches the LIN bus signal. Then, press down the knob to
select it. You can also press Version continuously or enable the touch
screen to select it. The available versions include 1.X, 2.X, and Both.
5. Display-related setting
Press Display to enter the display setting menu.
Set the display format
Press Format, rotate the multifunction knob
format of the bus data. Then, press down the knob to select it. You can also
press Format continuously or enable the touch screen to select it. The
11-24
(set
the
corresponding
or use the pop-up numeric keypad to set
Chapter 11 Protocol Decoding
LIN
decoding
parameters
to select the desired
to adjust the
to select a protocol
to select the display
MSO7000/DS7000 User Guide