RIGOL
MOSI).
Note:
When the source channel of MISO or MOSI is set to "OFF",
information of the data line is not displayed in the data table.
Open or close the event table: Press Event Table continuously to
enable or disable the event table. When you enable the event table,
the following figure is displayed, as shown in Figure 11-8. You can also
enable the touch screen to tap the icon
the table to close the event table. Besides, when the operating status is
"STOP", you can also rotate the multifunction knob
specified line and view the corresponding decoding information.
Line Number Time MISO Decoding Data MISO Error MOSI Decoding Data MOSI Error
Event Table
SPI Bus
Note:
Set the event table format: Press Format, rotate the multifunction
knob
the event table. Then, press down the knob to select it. You can also
press Format continuously or enable the touch screen to select it. The
available formats include "Hex", "Dec", "Bin", and "ASCII".
Set the view type: Press View, then rotate the multifunction knob
to select the form of the event table to be viewed. Press down the knob
to select it. You can also press View continuously or enable the touch
11-22
Figure 11-8 SPI Decoding Event Table
When you adjust the horizontal time base, the waveform displayed
on the screen will also change, and the total number of lines
containing the decoding information in the event table will also be
changed.
The displayed decoded data information in the bus is related to the
value of the horizontal time base. Reducing the horizontal time
base can help you view the detailed information.
to select the display format of "Data" (MISO and/or MOSI) in