Chapter 17 Function/Arbitrary Waveform Generator (MSO7000 Option)
continuously to select it. Also, you can enable the touch screen to select it.
Rotate the multifunction knob
offset or low level value of the current signal. When Impedance is set to "HighZ",
the available range is from (-2.5 V+currently set amplitude value/2) to (2.5 V-
currently set amplitude value/2); when Impedance is set to "50Ω", the available
range is from (-1.25 V+currently set amplitude value/2) to (1.25 V- currently set
amplitude value/2).
4. Set the start phase
Press Start Phase, then rotate the multifunction knob
numeric keypad to set the start phase of the current signal. Its available range is
from 0° to 360°.
5. Align phase
Pressing Align Phase will re-configure the two channels to output according to
the preset frequency and phase. For two signals whose frequencies are the same
or in multiple relationship, this operation can align their phases. Use the
oscilloscope to acquire the waveforms of the two channels and stably display the
waveforms. After switching the channel status, the phase deviation between the
two waveforms is changed. At this time, press Align Phase, and then the phase
deviation shown on the oscilloscope will restore to the current phase deviation
between the two waveforms automatically.
6. Setting options
Press Settings to enter the setting options menu. You can set the parameters
such as modulation, sweep, burst, and impedance.
Modulation, sweep, and burst
For details about modulation, sweep, and burst, refer to descriptions in
"Modulation", "Sweep", and "Burst".
Note:
When "Pulse", "DC", or "Noise" is selected under Wave, this menu is
not available.
Set the impedance
Press Impedance continuously to set the output impedance of the
Function/Arbitrary Waveform Generator. It can be set to "HighZ" or "50Ω".

To Output Square

Press Wave, rotate the multifunction knob
knob to select it. You can also press Wave continuously or enable the touch screen
to select it. At this time, you can set the parameters for the square waveform. For
setting methods, refer to descriptions in "To Output Sine". The duty cycle of the
square waveform is a fixed value 50%.
MSO7000/DS7000 User Guide
or use the pop-up numeric keypad to set the
to select "Square". Press down the
RIGOL
or use the pop-up
17-3