Chapter 8 Power Analysis (Option)
Press Mid, then rotate the multifunction knob
numeric keypad to set the middle value. The middle value is limited by the
settings of the upper limit and lower limit.
Press Lower, then rotate the multifunction knob
numeric keypad to set the lower limit.
Press Default, and then the upper value, middle value, and lower value will
be restored to the defaults.
Set Count
Press Count, then rotate the multifunction knob
keypad to set the number of statistical times for power quality analysis. Its
range is from 2 to 5,000. By default, it is 1,000.
Reset
Press Reset Stat. to clear the current data and execute statistics on the
measurement results again.
Open or Close the Results Display Window
Press Display continuously to enable or disable the display of the results.
When enabled, the statistical results will be displayed on the screen, as shown in
the figure below. You can also enable the touch screen to tap the icon
right-upper corner of the statistical results window to close it.
Note:
If you select "On" under Display, the multiplication operation will be
enabled in the Math operation.
View the Connection Diagram
Press Tips, and then the connection diagram of the power quality analysis is
displayed on the screen. Please connect the cables according to the connection
method as shown in Figure 8-1. Press Tips again or enable the touch screen to
tap the icon
close it.
MSO7000/DS7000 User Guide
at the right-upper corner of the connection diagram window to
or use the pop-up
or use the pop-up
or use the pop-up numeric
RIGOL
at the
8-3