Scale mA Channel to Temperature

The Product provides a function to convert mA current
channel reading to temperature reading.
To scale mA channel to temperature:
1.
Push .
2.
Push .
The temperature reading shows on the display.
To switch back to mA channel:
1.
Push .
2.
Push .
The mA current reading shows on the display.
Temperature at 4 mA = Span check 0 %
Temperature at 20 mA = Span check 100 %
Note
Scale mA Channel to Temperature

Set 0 % and 100 % Output Parameters

You must set the 0 % and 100 % points before you can
use the step and ramp functions or use mA/temperature
conversion. Proceed as follows:
1.
If necessary, push  for Source mode.
2.
Use the arrow keys to enter the value for 0 %.
3.
Push  and then  to set the 0 % value.
4.
Use the arrow keys to enter the value for 100 %.
5.
Push  and then  to set the 100 % value.
6.
Use , , , or  to adjust the value.
This function is only available when Source
mode is selected.
Thermocouple Calibrator
Note
21