8-6 DAR/ PI measurement, 1-min value display (KEW3552/3552BT only)

8-6-1 DAR/ PI measurement and 1-min value display

The instrument can measure and calculate DAR (Dielectric Absorption Ratio) 
and PI (Polarization Index) values automatically during an insulation resistance 
measurement. (not available on KEW3551)
●   T he  DAR  value  is  displayed  in  1  minute  and  the  PI  value  in  10  minutes 
after starting a measurement. 
● The following table shows the formula and display range.
Resistance value (1 min after a start of test)
DAR =       
Resistance value (15 sec after a start of test)
Formula
Resistance value (10 min after a start of test)
PI =       
Resistance value (1 min after a start of test)
Display range
*   I f  a  denominator,  the  measured  resistance,  used  in  the  above  formula  is  0 
MΩ;  the  LCD  shows no for  DAR/  PI  value.  When  the  calculated  DAR/PI 
values exceed the display range, the LCD shows >9.99 .

8-6-2 Indication

Press the cursor key, in 1 min. after starting a measurement, to check 1-min 
value or DAR/ PI values. (Please wait at least 10 min. to check PI value.)
The following indications show what value is currently displayed.
● 1-min value:
"
"symbol and the
measured value.
● DAR value:
"
symbol and the DAR value.
"
● PI value:
"
"
symbol and the PI value.