8-5-2 Reference value settings
To turn off this status indicator function or alter the reference value to any
desired value (available only on 3552/ 3552 BT), follow the procedures
described below.
* T he setting procedures depend on each model. Please refer to the
explanation about the model you re using.
[KEW3551]
(1) S et the function switch to the insulation resistance range you wish to set
the reference value.
(2) P ress the COMP button to toggle the values. The currently selected
settings can be checked on the LCD indications.
Fig. 8-13
Insulation status indicator is ON.
The LCD shows"COMP"and
the reference value.
(3) You can set this status indicator to ON or OFF for each range, respectively.
(The settings won t be cleared even after powering off the instrument.)
[KEW3552/ 3552BT]
(1) Get the instrument into setting mode.
1. S et the function switch to any position other than low-resistance
measurement, and then hold down the SETUP button ( 2 sec.) in
stand-by mode.
2. Use the cursor key and select the range to set the reference value.
"
"and the selected
range blink.
3. Press the ENTER button.
"
"and the reference
value start blinking.