8-5-2 Reference value settings

To turn off this status indicator function or alter the reference value to any 
desired value (available only on 3552/ 3552 BT), follow the procedures 
described below.
*   T he  setting  procedures  depend  on  each  model.  Please  refer  to  the 
explanation about the model you re using. 
[KEW3551]
(1)   S et  the  function  switch  to  the  insulation  resistance  range  you  wish  to  set 
the reference value.
(2)   P ress  the  COMP  button  to  toggle  the  values.  The  currently  selected 
settings can be checked on the LCD indications.
Fig. 8-13
Insulation status indicator is ON.
The LCD shows"COMP"and
the reference value.
(3) You can set this status indicator to ON or OFF for each range, respectively.
   (The settings won t be cleared even after powering off the instrument.)
[KEW3552/ 3552BT]
(1) Get the instrument into setting mode.
   1.   S et  the  function  switch  to  any  position  other  than  low-resistance 
measurement,  and  then  hold  down  the  SETUP  button  (   2  sec.)  in 
stand-by mode.
   2. Use the cursor key and select the range to set the reference value. 
 
"
"and the selected
range blink.
   3. Press the ENTER button.
"
"and the reference
value start blinking.
̶  27  ̶
www.calcert.com
15
10
20
5
25
0
30
Fig. 8-14
Insulation status indicator is OFF.
No"COMP"nor reference value
are displayed.
Fig. 8-15
Fig. 8-16