8-3 Voltage characteristics of measurement terminals

This  instrument  conforms  to  IEC61557.  This  standard  defines  that  the  rated 
measurement  current  shall  be  at  least  1  mA,  and  the  lower  limit  of  the 
insulation  resistance  maintaining  the  rated  measurement  voltage  at 
measurement  terminals.  (See  the  table  below.)  This  value  is  calculated  by 
dividing the rated voltage by rated current. ie., in case that the rated voltage is 
500 V, the lower limit of the insulation resistance is found as follows. 
 Divide 500 V by 1 mA equals 0.5 MΩ.
That  is,  insulation  resistance  of  0.5  MΩ  or  more  is  required  to  provide  the 
rated voltage to the instrument. 