8-5 Insulation status indicator
8-5-1 Reference value for insulation status, Pass/ Fail, judgement
The instrument can compare the measured insulation value and the pre-set
reference value listed below. LCD backlight lights up and buzzer sounds
according to the result. This function can be turned off.
With KEW 3552/ 3552BT, you can select any desired value as a reference
value; the reference values are fixed and unchangeable on KEW 3551 .
● Reference values for KEW3551
Rated voltage
50V
100V
Reference
0.1M
0.1M
value (Ω)
These reference values are determined based on the resistance value to
provide 1 mA with the rated voltage. The color of LCD backlight indicates the
result as follows.
Compared result
LCD backlight color
exceeding the reference value
Solid green
reference value or less
Solid red
Fig. 8-11
When a measured value is higher than
the reference value:"PASS"symbol
appears, and green LCD backlight lights
up.
̶ 26 ̶
www.calcert.com
15
10
20
5
25
0
30
125V
250V
500V
1000V
0.125M
0.25M
0.5M
1M
Fig. 8-12
When a measured value is equal to or
lower than the reference value:"FAIL"
symbol appears, and red LCD backlight
lights up.