Insulation Status Indicator; Reference Value For Insulation Status, Pass/ Fail, Judgement - KYORITSU KEW3552 Instruction Manual

Insulation - continuity tester
Hide thumbs Also See for KEW3552:
Table of Contents

Advertisement

1.888.610.7664

8-5 Insulation status indicator

8-5-1 Reference value for insulation status, Pass/ Fail, judgement

The  instrument  can  compare  the  measured  insulation  value  and  the  pre-set 
reference  value  listed  below.  LCD  backlight  lights  up  and  buzzer  sounds 
according to the result. This function can be turned off.
With KEW 3552/ 3552BT, you can select any desired value as a reference 
value; the reference values are fixed and unchangeable on KEW 3551 .
● Reference values for KEW3551
Rated voltage
50V
100V
Reference 
0.1M
0.1M
value (Ω)
These  reference  values  are  determined  based  on  the  resistance  value  to 
provide 1 mA with the rated voltage. The color of LCD backlight indicates the 
result as follows.
Compared result
LCD backlight color
exceeding the  reference value
Solid green
reference value or less
Solid red
 
Fig. 8-11
When a measured value is higher than
the reference value:"PASS"symbol
appears, and green LCD backlight lights
up.
̶  26  ̶
www.calcert.com
15
10
20
5
25
0
30
125V
250V
500V
1000V
0.125M
0.25M
0.5M
1M
Fig. 8-12
When a measured value is equal to or
lower than the reference value:"FAIL"
symbol appears, and red LCD backlight
lights up.
sales@calcert.com