Insulation - continuity tester
Hide thumbs Also See for KEW3552:
Table of Contents

Advertisement

1.888.610.7664
Source side
Earth terminal
Black
Red
Load side
Fig. 8-2
Indicates the elapsed time.
(6) Auto discharge function
This  function  allows  electric  charges  stored  in  the  capacitance  of  the  circuit 
under  test  to  be  automatically  discharged  after  measurement.  Set  the  test 
button or remote-control switch to off with the test leads connected. 
Discharge can be checked with the blinking
blinking LCD backlight.
# DANGER
Never touch the circuit under test immediately after measurement. 
Capacitances stored in the circuit may cause electric shock. Leave the test 
leads connected to the circuit, and do not touch the circuit until blinking 
goes off.
(7)   P ower  off  the  instrument  when  measurement  completes  and  disconnect 
the test leads. 
   ●   T he instrument holds the measured value display when a measurement 
completes. You can save the result in the internal memory in this state 
(3552/3552BT only). For the detail of memory function, please refer to 
13. Memory function.  The  held  display  can  be  released  by  turning  the 
function switch or starting another measurement.