4.3.1 Calibration accessories

CAUTION
×
The calibration plate is a certified measuring tool; hold by the base when fitting and removing.
×
Never touch the ceramic plate (test marks on the calibration plate could be damaged).
×
Do not drop the calibration plate (risk of breakage)
×
Never wipe the ceramic plate or clean it with an abrasive substance
(scratches on the plate may significantly impair the measurement result)

4.3.2 Preparation

×
Lay out the calibration plate, mounting plate and
4 spacer plates.
×
Start by clicking the
"Calibration"
×
Select the connected scanner
×
Confirm the selection by clicking "OK"
NOTE
The calibration procedure repeats itself 5 times at different heights.

4.3.3 Steps 1 to 5

1
step
2
step
st
nd
1
×
×
Insert the calibration
Insert first spacer
plate with the base
plate
plate
(#1)
in the
×
Confirm selection by
scanner.
clicking
×
Confirm selection by
clicking
"OK" 1
Calibration plate
white
button
1
2
3
step
rd
1
1
×
Insert first spacer
(#2)
plate
(#3)
×
Confirm selection by
"OK" 1
clicking
"OK" 1
4 × Spacer plates
1 × mounting plate
1
2
4
step
5
step
th
th
1
×
×
Insert first spacer
Insert first spacer
plate
(#4)
plate
×
×
Confirm selection by
Confirm selection by
clicking
"OK" 1
clicking
1
(#5)
"OK" 1
21