Introduction Purpose This manual is intended to assist the user in determining whether a reusable instrument has worn to an extent that it is no longer suitable for use. This document is provided in conjunction with the cleaning and sterilization instructions for reusable instruments. For additional reusable instrument instructions, see the Instructions For Use for Biomet 3i Kits and Instruments (Doc No.
Page 5
Note: Zimmer Biomet Dental does not specify the maximum number of uses appropriate for re-use. The useful life of the reusable instrument depends on many factors including the method and duration of each use as well as the handling between uses.
Bending Straight Osteotome Angled Osteotome Description Potential Effects of Wear • Bent • Inadequate force applied • Kinked • Unintended contact with soft tissue or bone • Crooked • Inaccurate feedback • Inadequate compression of bone • Incorrect angle...
Page 7
Description Potential Effects of Wear • Bent • Inadequate force applied • Kinked end • Unintended contact with soft tissue or bone • Crooked/ uneven end • Inaccurate feedback • Blunt / dull tip • Inadequate compression of bone / bone graft •...
Fracture Description Potential Effects of Wear • Fractured • Inadequate feedback • Broken • Inadequate force applied • Chipped • Bone fracture and/or splitting • Ingestion / Aspiration / Foreign body reaction...
Page 12
Unless otherwise indicated, as referenced herein, all trademarks are the property of Zimmer Biomet; and all products are manufactured by one or more of the dental subsidiaries of Zimmer Biomet Holdings, Inc. and marketed and distributed by Zimmer Biomet Dental and its authorized marketing partners.