6.8.6 Editing internal control parameters

Adjusting control parameters manually or
automatically

6.8.7 Editing external control parameters

Adjusting control parameters manually or
automatically
102 / 156
Press the [Parameter sets] button to display a list of available parameter sets.
If no data is available, the list will be empty.
When a data record is touched, it is displayed in a white font.
Pressing the [Display] button displays the selected parameter set. You
can read the data but you cannot modify it.
Pressing the [Load] button transfers the control parameters to the
thermostatic circulator automatically.
You can delete the selected parameter set by pressing the [Delete]
button and confirming a security prompt.
If you require the controller output limit, set it before adjusting the
control parameters. Ä Chapter 6.5 "Defining the actuating signal
limit" on page 89
The internal control in the device is active. Refer to Ä Chapter 6.6.1
"Activating external control, deactivating internal control" on page 89 for
information on reconfiguring the control.
The [Settings Tv/Td] menu item allows you to define whether the control
parameters [Tv] and [Td] are adapted manually or configured automatically.
If the automatic setting is enabled, both control parameters are marked with
the word "auto" and cannot be selected. In this case, [Tv] and [Td] are derived
with fixed factors of [Tn].
The slide switch in the [Settings Tv/Td] menu item allows you to define
whether the control parameters [Tv] and [Td] are adjusted manually or con-
figured automatically.
1.
Select the Control parameters  Internal menu item in the Control
menu.
2.
Touch a control parameter to select it.
An input window appears. A value can be entered within the limit
values displayed.
3.
Change the value accordingly.
4.
Press the [OK] button to return to the previous screen with the new
setting configured.
If you require the controller output limit, set it before adjusting the control
parameters. Ä Chapter 6.5 "Defining the actuating signal limit" on page 89
The external control in the device is active. Refer to Ä Chapter 6.6.1
"Activating external control, deactivating internal control" on page 89 for
information on reconfiguring the control.
The [Settings Tve/Tde] menu item allows you to define whether the control
parameters [Tve], [Tde] and [Prop_E] are adapted manually or configured
automatically. If the automatic setting is enabled, all three control parame-
ters are marked with the word "auto" and cannot be selected. In this case,
[Tve] and [Tde] are derived with fixed factors of [Tne].