Operating Elements; Mains And Safety Switch; Release Button And Maximum Temperature Knob; Functional Elements - Lauda P 10 Operation Manual

Hide thumbs Also See for P 10:
Table of Contents

Advertisement

3.2

Operating elements

3.2.1 Mains and safety switch

Fig. 6: Mains switch

3.2.2 Release button and maximum temperature knob

Fig. 7: Release button and Tmax knob
3.3

Functional elements

3.3.1 Hydraulic circuit

V6
The mains switch can be set to the following positions:
Position [I] switches the device on.
Position [O] switches the device off.
The mains switch also functions as a safety switch.
1.
If a fault current occurs, the fuse triggers and the mains switch moves
to position [0].
The device is switched off.
2.
Pull the mains plug from the power socket.
3.
Rectify the mains fault.
4.
Plug the mains cable into the power socket and set the mains switch
to position [1].
The device starts.
If the safety switch jumps back to the [O] position, contact LAUDA
Service Temperature control devices.
Set the maximum temperature T
Release button
Rotary knob with scale for setting the maximum permitted temperature
[Tmax], adjusted using a screwdriver. Refer to Ä Chapter 5.8 "Setting
overtemperature protection Tmax" on page 73 for more information on
setting the temperature.
In the event of a malfunction, press the [Release] button once the fault
is rectified.
The hydraulic circuit refers to the circuit that the heat transfer liquid flows
through.
PRO bath thermostats and circulation thermostats
max
21 / 156

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents