Lauda P 10 Operation Manual page 93

Hide thumbs Also See for P 10:
Table of Contents

Advertisement

Editing program examples
V6
Setting
Out 1, Out 2, Out 3
Delete
Undo delete
Copy
Graph
Fig. 55: Program example
The graph shows an example of a reprogrammed set temperature profile.
The cooling time in the graph varies depending on the device type, consumer
and so on. In example segment number 2, 50 °C should be reached within
20 minutes.
The original values in the first table below (before) are represented by a solid
line, while the values in the edited profile in the second table (after) are
represented by a dashed line.
PRO bath thermostats and circulation thermostats
Description
Switching contacts on the contact module (if avail-
able) can be programmed here (open, closed and
----). The entry ----- for contact modules indi-
cates that the previous segment was not modified.
If ----- appears in all the fields, this means that
the contact position of the initial setting or the set-
ting prior to program start is retained. The contact
modules are available as an accessory.
A segment (a program line) is deleted
A delete operation can be reversed
A segment is copied and inserted under the copied
segment
The program currently open is displayed as a tem-
perature-time diagram in the Cartesian coordinate
system
93 / 156

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents