Operation Status Register - R&S NRP Series User Manual

Hide thumbs Also See for NRP Series:
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
Figure 11-6: Standard event status register (ESR)
Table 11-8: Used standard event status bits and their meaning
Bit no.
0
2
3
4
5
6
7

11.2.5 Operation status register

Contains information on current operations, CONDition register, or operations per-
formed since the last query, EVENt register.
User Manual 1177.5079.02 ─ 15
Operation Complete
Query Error
Device-Dependent Error
Execution Error
Command Error
User Request
Power On
Short description
Operation complete
Query error
Device-dependent error
Execution error
Command error
User request
Power on
0
0
1
2
3
+
4
5
6
7
Bit is set if
All previous commands have been executed and
The controller wants to read data from the sensor but has not sent a
query, or it sends new commands to the sensor before it retrieves
existing requested data. A frequent cause is a faulty query which
cannot be executed.
A sensor-dependent error occurs. An error message with a number
between -300 and -399 or a positive error number denoting the error
in greater detail is entered in the error queue.
The syntax of a received command is correct but the command can-
not be executed due to various marginal conditions. An error mes-
sage with a number between -200 and -300 denoting the error in
greater detail is entered in the error queue.
An undefined command or a command with incorrect syntax is
received. An error message with a number between -100 and -200
denoting the error in greater detail is entered in the error queue.
The sensor is switched over to manual control.
The sensor is switched on.
Remote control basics
Status reporting system
is received.
*OPC