Selecting A Measurement Mode; [Sense<Sensor>:]Range:auto; [Sense<Sensor>:]Range:clevel - R&S NRP Series User Manual

Hide thumbs Also See for NRP Series:
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
Manual operation:
[SENSe<Sensor>:]RANGe:AUTO <state>
Enables or disables the automatic measurement path selection.
Parameters:
<state>
Manual operation:
[SENSe<Sensor>:]RANGe:CLEVel <level>
Reduces the transition range between the measurement paths, 0 -> 1 and 1 -> 2, by
the set value. Thus, you can improve the measurement accuracy for signals with a
high peak-to-average ratio, since the headroom for modulation peaks becomes larger.
However, the S/N ratio is reduced at the lower limits of the transition ranges.
Parameters:
<level>

9.4.3 Selecting a measurement mode

► Before starting a measurement, select the measurement mode using:
[SENSe<Sensor>:]FUNCtion
The available measurement modes and how to configure them are described in
Chapter 9.7, "Configuring the measurement
[SENSe<Sensor>:]FUNCtion <function>
Sets the measurement mode.
Parameters:
<function>
User Manual 1177.5079.02 ─ 15
0 is the most sensitive path.
2 is the most insensitive path.
1 is the path with medium sensitivity.
Range:
0 to 2
*RST:
2
See
"Range"
on page 50
*RST:
ON
See
"Range"
on page 50
Range:
-20.00 to 0.00
*RST:
0.00
Default unit: dB
"POWer:AVG"
Continuous average mode
See
Chapter 9.7.1, "Continuous average
on page 91.
"POWer:BURSt:AVG"
Burst average mode
See
Chapter 9.7.2, "Burst average
Remote control commands
Preparing for the measurement
modes",
on page 90.
measurement",
measurement",
on page 94.
71

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents