*OPC? basically functions like *WAI, but also returns a response. The response is an
advantage, because you can query the execution of commands from a controller pro-
gram before sending new commands. Thus preventing overflow of the input queue
when too many commands are sent that cannot be executed.

*OPT?

Option identification
Returns a comma-separated list of installed options.
Usage:
*PRE <register>
Parallel poll register enable
Sets the parallel poll enable register to the specified value or queries the current value.
Parameters:
<register>
*RCL <number>
Recall
Calls the device state which has been stored with the
fied number.
Setting parameters:
<number>
Usage:

*RST

Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command corresponds to the
Usage:
*SAV <number>
Save
Stores the current device state under the specified number.
User Manual 1177.5079.02 ─ 15
Query only
Range:
0 to 255
*RST:
0
Range:
0 to 9
*RST:
0
Setting only
SYSTem:PRESet
Event
Remote control commands
Common commands
command under the speci-

*SAV

command.