9.4 Preparing for the measurement

Before starting a measurement, you need to do the following:

9.4.1 Selecting the reference source

The ROSCillator subsystem contains commands for configuring the reference
source.
[SENSe<Sensor>:]ROSCillator:SOURce <source>
Sets the source of the reference oscillator. Refers typically to a precision, stabilized
time base.
<source>
Example:

9.4.2 Selecting a measurement path

The RANGe subsystem contains commands for selection of a measurement path.
[SENSe<Sensor>:]RANGe <range>
Sets the selected path as active measurement path.
Parameters:
<range>
User Manual 1177.5079.02 ─ 15
Event
source............................................................................... 70
path............................................................................... 70
mode............................................................................. 71
values........................................................................... 72
INTernal | EXTernal | HOST
INTernal
Internal precision oscillator
EXTernal | HOST
External signal supplied at the host interface connector.
If the sensor boots or reboots, the source is set to
INTernal. If the sensor is reset, the source setting is
kept unchanged.
ROSC:SOUR INT
................................................................................................70
......................................................................................71
.................................................................................... 71
The sensitivity of the paths differs.