Braun Aesculap S4 MIS Instructions For Use/Technical Description page 8

Instruments
Hide thumbs Also See for Aesculap S4 MIS:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
®
Aesculap
S4 MIS instruments
Excessive trocar insertion
depth in the event of a missing
end stop means a risk of
WARNING
injury!
Delay in surgery due to the
outer sleeve slipping down
FW705R!
WARNING
Damage to the pedicle screws
and incorrect positioning due
to using the wrong screw-
CAUTION
driver!
4
Use trocar FW271R only in
combination with the aim-
ing device FW258R.
Secure the outer sleeves by
hand when inserting or
removing other instru-
ments (e.g. the thread cut-
ter, screwdriver).
The outer sleeve can be
reapplied using the posi-
tioning aid FW708R.
The outer sleeve can be
secured using the threaded
tube FW234R and the
adjusting screw, long
FW707R
Only use the guide instru-
ment FW708R for screw-
ing in monoaxial pedicle
screws.
Damage to the monoaxial ped-
icle screw when screwing in
with the FW708R with the
CAUTION
screwdriver in the wrong posi-
tion!
Disassembling
The threaded tube FW734R and adjust-
ing screw FW707R
When using the adjusting screw FW707R with
4
S
percutaneous threaded tube FW734R: Screw the
adjusting screw 4 off the threaded tube FW734R,
see Fig. A.
Spacer FW709R
Remove the nut and the guide sleeve a from the
spacer 5, see Fig. B.
Bar insertion instrument FW711R
Screw the inner part out of the bar insertion
instrument 6, see Fig. C.
Distractor FW238R and distraction
blades, short FW706R
Activate slider b and remove the distraction
blades 7 from the distractor FW238R, see Fig. D.
Dismantle distractor FW238R with the distraction
blades.
When screwing in the
monoaxial pedicle screw,
always insert the screw-
driver all the way into the
head of the pedicle screw.