Aesculap
®
S4 MIS instruments
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with
immersion disinfection
FW705R–FW709R
FW711R
FW712R
FW714R–FW716R
FW734R
Mechanical alkaline
cleaning and thermal
disinfection
FW712R
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
FW705R–FW709R
FW711R
FW714R–FW716R
FW734R
8
Specific requirements
Suitable cleaning brush
20 ml disposable syringe
Keep working ends open for clean-
ing.
When cleaning products with mov-
able hinges, ensure that these are in
an open position and, if applicable,
move the joint while cleaning.
Drying phase: Use a lint-free cloth
or medical compressed air
Place the product in a tray that is
suitable for cleaning (avoiding rins-
ing blind spots).
Connect components with lumens
and channels directly to the rinsing
port of the injector carriage.
Keep working ends open for clean-
ing.
Place products in the tray with their
hinges open.
Suitable cleaning brush
20 ml disposable syringe
Place the product in a tray that is
suitable for cleaning (avoiding rins-
ing blind spots).
FW705R/FW709R/FW714R–
FW716R: Connect components with
lumens and channels directly to the
rinsing port of the injector carriage.
Keep working ends open for clean-
ing.
Place products in the tray with their
hinges open.
Reference
Chapter Manual cleaning/disinfection
and sub-chapter:
Chapter Manual cleaning with
immersion disinfection
Chapter Mechanical cleaning/disinfect-
ing and sub-chapter:
Chapter Mechanical alkaline clean-
ing and thermal disinfecting
Chapter Mechanical cleaning/disinfec-
tion with manual pre-cleaning and sub-
chapter:
Chapter Manual pre-cleaning with a
brush
Chapter Mechanical alkaline clean-
ing and thermal disinfecting