Liquid Sanitization - ELGA PURELAB Classic - US Operator's Manual

Table of Contents

Advertisement

PURELAB Classic Operator Manual
Screens
Page 38
8.2

Liquid sanitization

The unit is sanitized to reduce the growth of microbiological
contamination within the unit. Please read this entire section to
become familiar with the procedure before you start. Sanitization may
be required in the following circumstances:
 To maintain low bacterial counts
 If the unit has not been used for a prolonged period of time
 If the product water is used for particularly stringent
applications
Step 1 – Start Sanitization Cycle
1. ENSURE unit is ready to use.
2. SWITCH OFF electrical supply.
3. REMOVE POU filter, if fitted.
4. OPEN the door.
Step 3 – Add Minncare Cold Sterilant into LA698-US Liquid
Sanitization Block
1. UNSCREW
sanitization block.
2. ADD 10ml of Minncare Cold Sterilant into the container.
There is a line on the container to assist you.
3. REASSEMBLE
container back onto the main body.
CAUTION!
4. REFIT the container on to by-pass block, hand tight.
PURELAB Classic Version 2 11/15
the
container
from
the
LA698-US
by
ENSURE THE CONTAINER REMAINS
UPRIGHT SO THAT THE CHEMICAL
REMAINS WITHIN THE LA698-US.
ELGA
the
LA698-US
screwing
the