ELGA
Fit By-pass Block
Slide out
Tilt
Push
Hangers
Remove By-pass Block
Step 4 - Fit By-pass Block
Spigots
Bypass block
Step 5 - Remove By-pass Block
PURELAB Classic Version 2 11/15
1. WET 'O' rings on by-pass block.
2. SLIDE by-pass block into unit.
3. POSITION by-pass block onto appropriate spigots.
4. ENSURE by-pass is locked in place.
5. Connect temporary tube from the dispense outlet to a
suitable drain or sink.
6. SWITCH ON power. The screen will display the insert
tablet icon, press
7. The unit will display the sanitization recirculation icon to
indicate that the unit is ready to commence the
sanitization procedure. Press
cancel sanitization procedure.
8. Upon selecting to continue the sanitization procedure
the unit will enter a 10 minute recirculation cycle.
9. On completion of this cycle the DI and UV versions will
automatically start a 44 minute rinse sequence. The UF
and UVF versions will display the UF "flush valve open"
icon indicating that the manual UF rinse valve must
be opened.
10. Press the dispense knob to continue and the unit will
commence the flush sequence.
11. The UF and UVF versions will dispense to drain for 3
minutes (countdown indicated on the display) the unit
will then sound a buzzer and stop. The UF "flush valve
close"
icon will appear indicating the manual UF
flush valve should be closed. CLOSE the UF flush
Valve and press the dispense knob to continue. The
unit will then start a 44 minute rinse sequence until
completion.
12. The UF and UVF versions will display the UF "flush
valve open"
icon indicating that the manual UF rinse
valve must be opened. Open the valve and press the
dispense knob to continue. The unit will commence the
flush sequence. The UF and UVF versions will dispense
to drain for 3 minutes (countdown is indicated on the
display), the unit will then sound a buzzer and stop. The
UF "flush valve close"
manual UF flush valve should be closed.
13. Disconnect the temporary drain line.
14. On completion of the rinse the sanitization cycle is now
finished. The unit will now display the next sanitization
reminder screen. To set the sanitization reminder alarm
press
or
to cancel the next reminder.
1. PUSH and TILT by-pass block.
2. SLIDE out of unit.
3. PLACE on hanger located inside the door.