Cleaning Inlet Strainer - ELGA PURELAB Classic - US Operator's Manual

Table of Contents

Advertisement

ELGA
Inlet Strainer
The Inlet Strainer should be checked and cleaned every six months to
ensure that the strainer does not become clogged.
7.5

Cleaning Inlet Strainer

Step 1 - Remove the Inlet Strainer
1. SWITCH OFF electrical supply.
2. ISOLATE inlet water supply.
3. HOLD inlet strainer over a sink or receptacle.
4. DEPRESS collars on both sides of strainer and
disconnect tubing.
5. REMOVE the inlet strainer from its position.
Step 2 - Dismantle the Inlet Strainer
1. HOLD inlet strainer over a sink or receptacle.
2. UNSCREW inlet strainer.
3. REMOVE mesh filter.
4. CHECK mesh filter for signs of wear or damage,
replace or clean as necessary.
Step 3 - Reassemble the Inlet Strainer
1. INSERT mesh filter into strainer, ENSURE it is facing
the correct direction (see diagram).
2. RE-ASSEMBLE the inlet strainer.
Step 4 - Replace the Inlet Strainer
1. REPOSITION the inlet strainer.
2. REFIT tubes to inlet strainer, ENSURE it is facing the
correct direction.
3. RE-ESTABLISH inlet water supply.
4. SWITCH ON power.
PURELAB Classic Version 2 11/15
PURELAB Classic Operator Manual
Page 33