ELGA PURELAB Classic - US Operator's Manual page 35

Table of Contents

Advertisement

PURELAB Classic Operator Manual
Page 30
Step 7 - Replace the UV
1. See Section 7.4 (Step 5).
Step 8 - Rinse New UF
CAUTION!
1. SWITCH ON power. The unit will enter sanitization
mode. The insert tablet icon will appear, PRESS TICK
to continue.The sanitization recirculation icon will
appear, press tick to continue. The unit will begin a 10
minute recirculation.
2. Once the cycle has been completed, the unit will display
"open flush valve"
valve should be opened.
3. CONNECT a temporary tube from the dispense outlet
to a suitable drain or sink.
4. OPEN the flush valve and press the dispense knob, the
system will begin its flush sequence.
5. After approximately 3 minutes (countdown is indicated
on the display), the unit will sound a buzzer and stop.
The unit will display "close flush valve"
the flush valve should be closed.
6. CLOSE the flush valve and press the dispense knob,
the system will begin its rinse sequence.
7. Upon completion of the rinse sequence, the "open flush
valve"
and PRESS the dispense knob to start the 3:00 flush
sequence again.
8. Once this has been completed, the "close flush valve"
icon will appear again. CLOSE the flush valve and
PRESS the dispense knob to continue. The flush is
then complete after a 5 second flush to drain.
9. Ignore the prompt to reset the sanitization date, PRESS
TICK
button.
Step 9 - Set UF Change Reminder
1. See Section 5.5 - Setting Up Password/Replacement
Timers.
PURELAB Classic Version 2 11/15
The new Ultra Filter must be rinsed. To
rinse the unit enter the Sanitizing
Procedure. Ensure the by-pass block is
correctly fitted.
indicating that the manual flush
icon will appear again. OPEN the flush valve
ELGA
indicating that