ELGA PURELAB Classic - US Operator's Manual page 24

Table of Contents

Advertisement

ELGA
Water Purity Unit Setting Screen
Uncompensated Water Quality Screen
Outlet Purity Alarm Setting Screen
Step 6 - Water Purity Unit Setting
This screen allows preferred units of water purity to be set, to
either, M.cm or µS/cm.
1. PRESS SCROLL
2. PRESS TICK
Step 7 - Uncompensated Water Quality
A U will indicate uncompensated readings in the normal
process screen.
1. PRESS SCROLL
uncompensated reading is required.
2. PRESS TICK
Note:
The on-going display of uncompensated
values is generally not recommended as it
can lead to confusion among users and
increase the possibilities of dispensing water
of inadequate purity.
Step 8 - Product Water Purity Alarm Settings
This screen is used for setting the purity value at which the
product water purity alarm will activate. The unit will alarm if this
level is passed but will not switch off the process. The alarm will
automatically clear if the water purity returns above its specified
limit.
1. PRESS SCROLL
2. PRESS TICK
PURELAB Classic Version 2 11/15
PURELAB Classic Operator Manual
button to highlight appropriate box.
button.
button to highlight box if
button.
button to highlight appropriate box.
button.
Page 19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents