Parameter
Source - Originally from potable supply,
then pre-treated
Fouling Index (max)
Service Deionisation (SDI) - M-cm
Reverse Osmosis (RO) - µS/cm
Free Chlorine
TOC
Carbon Dioxide
Silica
Particulates
Temperature
Flowrate (maximum requirement)
Drain requirements (gravity fall with air gap).
Maximum during service
Feedwater Pressure
* Contact technical support for advice on feedwaters outside the range listed.
Note:
Different system configurations are available for different feedwater sources. See system set up.
Height
Width
Depth
Weight
DI
UV
UF
UVF
Inlet-quick connect
Flush - UF/UVF
Positioning
Environment
3.3

Technical Specification

The Technical Specifications for the PURELAB Classic are as
follows:
*
Feedwater
Limits
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.
Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion
1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds.
1 M-cm minimum resistivity at exhaustion.
Recommended < 30 µS/cm
0.05 ppm max.
Recommended 50 ppb max.
30 ppm max.
2 ppm max.
Filtration down to 0.2 micron advisable to protect internal and/or point of use filters.
1 - 40°C - Recommended 10 - 15°C
130 l/hr
Up to 2 l/min
0.7 bar (10 psi), maximum, 0.07 bar (1 psi) minimum
Dimensions
490 mm (19.3")
410 mm (16.2")
365 mm (14.4")
14.0 kg (30.8 lb)
14.5 kg (32.0 lb)
14.5 kg (32.0 lb)
15.0 kg (33.1 lb)
Connections
8 mm (5/16") OD
8 mm (5/16") OD
Wall, bench or under bench mounted.
Clean dry indoor. Temp 5 - 40°C.
Humidity max 80% non-condensing.
PURELAB Classic Version 2 11/15
PURELAB Classic Operator Manual
.
Page 9