5. INSTALLATION INSTRUCTIONS

5.1

Unpacking the PURELAB Classic

The following items are supplied with your PURELAB Classic:
1. PURELAB Classic unit.
2. By-pass block LA638 (1 off, fitted in unit).
3. 6 metres of 8mm (5/16") O/D tube.
4. 1 off Strainer Assembly.
5. Operator Manual in English.
6. Mains Lead.
7. Reservoir Level Connecting Lead.
5.2

Positioning the PURELAB Classic

Before installation and operation of the PURELAB Classic unit,
please read and observe the following points.
Environment
The unit should be installed on a flat, level surface, in a clean,
dry environment. The unit can also be wall mounted against a
vertical wall capable of supporting the weight (for this we
recommend the use of the wall mounting kit Part No. LA610
on substantial brick/concrete walls or LA622 for stud partition
walls).
WARNING
!
Note:
The unit is designed to operate safely under the following conditions:
 Indoor Use
 Altitude up to 2000 m
 Temperature Range 5 - 40°C
 Maximum Relative Humidity 80% @ 31°C decreasing
linearly to 50% @ 40°C, non-condensing
The unit is in Installation Category II, Pollution Degree 2, as per
IEC1010-1.
PURELAB Classic Version 2 11/15
PURELAB Classic Operator Manual
IF THE UNIT IS TO BE WALL MOUNTED,
ENSURE IT IS MOUNTED USING THE
CORRECT WALL MOUNTING KIT AND
THAT THE WALL IS CAPABLE OF
SUPPORTING THE OPERATING WEIGHT
OF THE SYSTEM. ALWAYS CAREFULLY
FOLLOW
THE
INCLUDED IN THE KIT.
Refer to Specifications for unit weights
(Section 3.3 - Technical Specification).