ELGA PURELAB Classic - US Operator's Manual page 16

Table of Contents

Advertisement

ELGA
Power fail safe
Water temperature alarm
Water quality alarm
Purification pack interlock
UV current monitoring
Timeout of dispense
Low operating voltage 24 V
Volt free contact alarm connection
Visual alarms
Audible alarms
Low noise levels – minimum intrusion
Variable dispense flowrate
Restart on power interrupt
Optional printer kit to record operating parameters
Optional remote display
Flowrate
Inorganic
TOC
Bacteria
Pyrogens
pH
Particles
RNase / DNase
Capacity
(LC162)
Confirms to ASTM, CAP, NCCLS – Type 1 and BS3978 Grade 1 Specification.
Safety Features
Operational Features
Purified Water Specification
DI
2.0 l/min max.
Resistivity 18.2 M.cm (0.055S/cm)
**<15 ppb
**<5 ppb
***<1 cfu/ml
<1 cfu/ml
***0.2 m
***0.2 m
-
45,000 liters
>18M.cm per single purification pack/S at pH 7.0
70,000 liters
>1M.cm per single purification pack/S at pH 7.0
*
On some models only.
**
Dependant on feedwater - recommended RO feed <50 ppb TOC.
*** With POU filter fitted.
As part of our policy of continual improvement we reserve the right to
alter the specifications given in this document.
PURELAB Classic Version 2 11/15
PURELAB Classic Operator Manual
UV
UF
**<15 ppb
<1 cfu/ml
<0.02 Eu/ml
Effectively neutral
Ultrafiltration
UVF
**<10 ppb
<1 cfu/ml
<0.005 Eu/ml
Page 11