Once the PURELAB Classic unit has been positioned either on a wall
or on a bench, it should be connected as follows:
Pull plug
Tube 8mm OD
PURELAB Classic Version 2 11/15

Connecting the PURELAB Classic

 Mains water inlet tube
 UF drain tube - to drain
Step 1 - Remove Transit Plugs
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8 mm OD semi-rigid
drain tube.
4. PUSH tube into connector.
Step 2 - Connect Water Inlet
1. Connect 8 mm tube from the PURELAB Classic inlet to
the inlet strainer.
2. Connect the inlet strainer to the pre-purified water
supply.
Step 3 - Connect Drain
1. Connect 8 mm tube from the UF drain (if fitted) and
direct to a suitable drainage point.
CAUTION!
The drain line should allow a gravity
fall to drain with no restrictions.
PURELAB Classic Operator Manual
Page 15