PURELAB Classic Operator Manual
Page 40
Step 8 - Return to Normal Operation
1. PRESS PROCESS
The unit has now been sanitized but should be allowed
to rinse to a suitable quality. For particularly sensitive
applications
overnight. To facilitate a quicker system recovery it is
suggested to dispense a volume of approximately 10
litres.
Note:
Once a sanitization cycle has commenced the
PURELAB Classic cannot be used to purify
water until the cycle has been totally
completed. If the electricity supply to the
PURELAB Classic is interrupted whilst the
cycle is in process, when it is restarted the
unit will resume sanitization at point of
interruption.
After sanitization the PURELAB Classic will
default to continuous recirculation for a
twelve-hour period.