Automatic Printout After Each Shock (Autoprint); Printout Of The Result Memory - Primedic DefiMonitor XD Series Operating Manual

Table of Contents

Advertisement

6.5.2
6.5.3
21324 / 09.05 / 02 / GB
DefiMonitor XD
All relevant parameters are printed out in a data paragraph:
Date, time
Energy (Joule)
Synchronised / asynchronised mode
Derivation
Heart frequency
Signal reinforcement
Lines for later notes (e.g. patient note etc.)
SpO
-value (only model XD 3, 30, 300 and 330)
2
Speed
In the device versions with SpO
curve in the log printout.
The printout is at a 5 second delay to the monitor display, i.e. events before the acti-
vation of the printout can be shown. If the printput is stopped, the printout ends 5
seconds in the past.
To separate the EKG log strip, use the integrated tear-off edge on the printer cover.
Detathc the papert strip laterally in an upward motion.
Systematic and brief printing saves energy and paper and extends the operating time
in which the device can work independently of the power supply. This particularly
applies to printing with 50 mm feed speed.

Automatic Printout after Each Shock (Autoprint)

The PRIMEDIC™ Defimonitor offers the ability to log the event automatically accord-
ing to defibrillation / cardio version. Here the results are documented 5 seconds be-
fore and 5 seconds after the shock.
The Autoprint function can be switched on and off in the setup menu. The function is
switched off in delivery state.
If the Autoprint function was switched on in the setup menu, this stays active, even
after shutting off the defibrillator or changing the battery. To deactivate this function,
you must change the device configuration in the setup menu.

Printout of the Result Memory

The PRIMEDIC™ DefiMonitor saves the last 30 defibrillations / cardio versions /
events automatically in an event memory. Here the EKG is saved (5 seconds before
and 5 seconds after the shock), the pulse curve (only at XD 3, 30, 300 or 330 if SPO
is active, as well as all relevant parameters.
A newly added result "pushes" the oldest result from the memory.
The memory content, beginning with the last recorded event, is print out from the
setup menu.
Go to the setup menu and select the menu item "Memo Print" with the cursor keys.
Confirm the point "activate". The memo-print starts. To stop the printout confirm
with the printer on/off key. The printout is at 25 mm/s.
The data in the result memory also remain after the printout. They can be printed out
any number of times.