Inserting The Power Module - Primedic DefiMonitor XD Series Operating Manual

Table of Contents

Advertisement

Operating Manual
5.3.1
26

Inserting The Power Module

1.
Fig. 15
PRIMEDIC™ DefiMonitor – Inserting the Power Module
Place the device on its rear side.
Push the (new) power module (2) in the direction of the arrow (1.) into the device
until it fits at the back (3) as shown in the illustration.
Then press the power module in the front in the direction of the arrow (2.) into the
compartment (1) until the release button (4) the tab of the power module are locked
securely.
Push in the power module into the device completely until you hear the 'click' of the
locking and the power module is locked evenly on the outer side of the device. If the
power module falls out when the device is moved, it has not been locked correctly.
If the power module was placed correctly, the device switches on and carries out an
automatic self test. Follow the acoustic / visual prompts of the device and switch the
device off afterwards.
Watch the Status-display. If the display shows "OK", the device is ready to be used
Switch the device (if necessary) off by pressing the On/Off button.
Does the display not show "OK", resp. an error message appears on the Monitor,
remove the cause or call your next service partner. The device switches off itself.