Device Self-Test; Self-Test After Switching On The Heartsave Pad; Automatic, Periodic Self-Tests - Primedic HeartSave PAD Operating Instructions Manual

Table of Contents

Advertisement

Device self-test

6 Device self-test

6.1 Self-test after switching on the HeartSave PAD

The self-test is triggered by switching on the PRIMEDIC™ HeartSave PAD or by
inserting the battery in the device. The HeartSave PAD runs a device self-test to check
all the important parameters and signal mechanisms.
If the power module has been changed and if the device had previously found a fault,
then the full self-test (FULL) is automatically triggered. Please proceed to follow the
device instructions.

6.2 Automatic, periodic self-tests

The Primedic HeartSave PAD automatically carries out self-tests to ensure that it is
ready to use.
Frequency
Daily
SHORT
First day of the month
MEDIUM
st
On July 1
and January 1
LONG
every year
After changing the power
FULL
module or detection of an
internal fault
Operating instructions
HeartSave PAD 22289 / GB / C
Effect of test
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 0 V
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 300
V
st
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 1600
V
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 1600
V, microphone test
27

Advertisement

Table of Contents
loading

Table of Contents